M2RTS-R-BF-R2SLASHCE
AI

The model **M2RTS-R-BF-R2/CE** belongs to the **M2 Series** of signal conditioners manufactured by **M-System**. This specific device is a thermocouple transmitter designed to convert temperature signals into standard process signals.
---
### 1. Technical Specifications Breakdown
The alphanumeric code describes the specific hardware configuration of the module:
| Component | Code | Description |
| :--- | :--- | :--- |
| **Model** | M2RTS | Thermocouple Transmitter (Linearization type) |
| **Configuration** | R | Removable terminal block |
| **Input** | BF | Universal Input (Field-configurable for Types K, E, J, T, B, R, S, N) |
| **Output** | R2 | 4 – 20 mA DC (Current output) |
| **Power Input** | SLASH | Separator |
| **Power Supply** | CE | 100 – 240V AC |
---
### 2. Key Electronic Components & Features
#### A. Signal Input Section
* **Cold Junction Compensation (CJC):** The unit includes an internal thermistor to compensate for the ambient temperature at the terminal block, ensuring accurate thermocouple readings.
* **Burnout Protection:** It features an electronic circuit that detects an open circuit (broken wire) in the thermocouple and drives the output to a pre-set level (usually upscale).
#### B. Isolation Barrier
The "M2" series is known for its **three-way isolation**:
1. **Input to Output**
2. **Input to Power**
3. **Output to Power**
*This prevents ground loops and protects sensitive control equipment (like a PLC) from high-voltage surges on the field side.*
#### C. Power Supply (CE Rating)
* **Switch-Mode Power Supply:** The "CE" designation indicates a wide-range AC power input (100-240V). It uses internal rectification and regulation to provide stable DC voltage to the internal sensing circuitry.
#### D. Output Stage
* **Current Loop (R2):** The output utilizes a precision operational amplifier and a transistor stage to provide a stable 4-20mA signal, capable of driving loads up to 550 $\Omega$.
---
### 3. Installation Summary
* **Mounting:** DIN rail mounting.
* **Terminals:** M3 screw terminals (removable for easy maintenance).
* **Adjustments:** Front-facing Zero and Span adjustments (potentiometers) for fine-tuning the calibration.
- ⤷
How do I configure the burnout protection (Upscale vs Downscale) on this model?
- ⤷ What is the maximum load resistance for the 4-20mA output?
- ⤷ Does this model support Modbus or digital communication?