  407227 Datasheet HTML 1Page - Extech Instruments Corporation. 407227 Datasheet HTML 2Page - Extech Instruments Corporation. 407227 Datasheet HTML 3Page - Extech Instruments Corporation. 407227 Datasheet HTML 4Page - Extech Instruments Corporation. 407227 Datasheet HTML 5Page - Extech Instruments Corporation. 407227 Datasheet HTML 6Page - Extech Instruments Corporation.  
Zoom Inzoom in Zoom Outzoom out
 3 / 6 page
background image
Model 407227 Ver. 2.1 Oct. 2002
Page 3
Temperature Compensation
For accurate pH measurements and calibration, the temperature of the solution under test
must be known. This can be accomplished manually or automatically. Read through the steps
for each (below) and determine that best method for the application at hand:
Automatic Temperature Compensation
NOTE: For Automatic Compensation, the supplied thermometer must be used. Perform
these steps before each use or when the temperature of the solution under test changes.
The thermometer can remain in the solution while taking pH measurements in order to
track temperature changes and compensate for them if necessary.
1. Select ‘pH ATC’ on the meter’s front panel slide-switch.
2. Plug the supplied thermometer into the meter’s temperature sensor input jack.
3. Place the other end of the thermometer into the solution under test.
4. Perform a pH Calibration or take a pH measurement as described later in the manual.
Manual Temperature Compensation
NOTE: For Manual Compensation, an external thermometer is required (separate from the
one supplied with the meter). Perform these steps before each use or when the
temperature of the solution under test changes
1. Place a thermometer in the solution under test and note the reading.
2. Select ‘pH MTC’ on the meter’s front panel slide-switch.
3. Press the "MTC" key and 25.0
oC will display on the LCD temperature display line.
Using the ">>" key (for 1
oC steps) and the ">" key (for 0.1oC steps), edit the displayed
temperature. Continue editing until the displayed temperature equals the measured
temperature that was noted earlier.
4. Perform pH calibration or pH measurement normally.
Calibrating a new thermometer
1. Connect the thermometer to the meter’s temperature sensor input jack.
2. Select "pH ATC" via the meter’s front panel slide-switch.
3. Place the temperature probe and a mercury thermometer in water.
4. Adjust VR6 (inside the battery compartment, see diagram below) until the displayed
value matches the mercury thermometer reading.
VR6
Inside battery compartment



Html Pages

1 2 3 4 5 6


数据表 下载

Go To PDF Page


链接网址



ALLDATASHEET是否为您带来帮助?  [ DONATE ] 

关于 Alldatasheet   |   广告服务   |   联系我们   |   隐私政策   |   数据表链接    |   链接交换   |   制造商名单
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com