9 / 18 page ![]() SA9603C sames 9/18 From the two bytes of data output by the device, the value for the register can be derived as shown: Register value = (First Byte * 256) + Second Byte The register value is represented in two's compliment in order to provide for positive and negative register values. The most significant bit of the 16 bit energy register (active or reactive) may be used as an indication of the direction of the energy flow (0 = positive, 1 = negative). The register access codes which may be written to the SA9603C via the serial communications port, are shown in the table below: REGISTER READ RESET READ-RESET DUMP ACTIVE $01 $02 $03 $05# RE-ACTIVE $81 $82 $83 $84 FREQUENCY $41 - - $44 VOLTAGE $C1 $C2 $C3 $C4 #Note: The Dump Active ($04) and Read Active ($01) commands have the same effect. Data on SOUT Commands for the voltage integrator D R - 00835 FI R S T B Y TE 8 9 10 11 12 13 14 15 S E C O ND BY T E 0 1 2 3 4 56 7 DUM P STA RT BIT DR-01362 RD / RE S DUM P RD RE S RD RE S RD RE S STO P BIT |
|
|
链接网址 |
| ALLDATASHEET是否为您带来帮助? [ DONATE ] |
关于 Alldatasheet | 广告服务 | 联系我们 | 隐私政策 | 数据表链接 | 链接交换 | 制造商名单 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|