22 / 29 page ![]() 22 PROBLEM POTENTIAL CAUSES SOLUTIONS IntelliTune fails because of a “load short” error VLIM may be too high. Reduce the current limit and rerun IntelliTune to automatically reduce VLIM. Then reset the current limit and rerun the scan. Use remote command TEC:VLIM 9 to set the VLIM to 9 V and then rerun IntelliTune. Put a resistor in series with the thermoelectric to increase the load compliance voltage. Load Fault Error is triggered A Load Fault Error may be falsely triggered when operating near maximum current and near maximum voltage. Use remote command TEC:VLIM 27 to limit the internal voltage supply to 27 V instead of the 28 V maximum. |
|
|
链接网址 |
| ALLDATASHEET是否为您带来帮助? [ DONATE ] |
关于 Alldatasheet | 广告服务 | 联系我们 | 隐私政策 | 数据表链接 | 链接交换 | 制造商名单 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|