29 / 30 page ![]() AD8233 Data Sheet Rev. 0 | Page 28 of 29 SYNCHRONIZED ECG AND PPG MEASUREMENT In wearable devices developed for monitoring the health care of patients, it is often necessary to have synchronized measurements of biomedical signals. For example, a synchronous measurement of a ECG and photoplethysmograph (PPG) can be used to determine the pulse wave transit time (PWTT), which can then be used to estimate blood pressure. The circuit shown in Figure 77 shows a synchronous ECG and PPG measurement using the AD8233 and the ADPD105 photometric front end. The AD8233 implements a two-pole, high-pass filter with a cutoff frequency of 0.3 Hz, and a two-pole, low-pass filter with a cutoff frequency of 37 Hz. The output of the AD8233 is fed to one of the current inputs of the ADPD105 through a 50 kΩ resistor to convert the voltage output of the AD8233 into a current. The PPG signal is acquired by the ADPD105, which is a complete optical transceiver with integrated LED drivers, multiple photodiode current inputs, an integrated, 14-bit, successive approximation (SAR) ADC, and a FIFO. In the circuit shown, the chip scale ADPD105 is used; the ADPD105 is a two input device. The ADPD105 is configured to alternately measure the photodiode signal and the ECG signal from the AD8233 on consecutive time slots to provide fully synchronized PPG and ECG measurements. Data can be read out of the on-chip FIFO or straight from the data registers. The ADPD105 channel that processes the ECG signal must be set up in either pulse connect mode or transimpedance amplifier (TIA) ADC mode, and the input bias voltage must be set to the 0.9 V setting. The TIA gain setting can be set to optimize the dynamic range of the signal path. The channel used to process the PPG signal is configured in its normal operating mode. Figure 76 shows a plot of a synchronized ECG and PPG measurement using the AD8233 with the ADPD105. 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 TIME (Seconds) SHOWN WITH fSAMPLE = 100Hz PPG ECG Figure 76. Synchronous ECG and PPG Measurement Using the AD8233 with the ADPD105 RL RA LA 1.8V 1.8V +IN –IN HPDRIVE +VS HPSENSE IAOUT REFIN GND FR AC/DC SDN LOD RLD RLDFB OUT OPAMP+ OPAMP– REFOUT SW AD8233 6.8nF 1MΩ 1MΩ 250kΩ 2.7nF 100kΩ 180kΩ 180kΩ 1MΩ 10MΩ 10MΩ 0.1µF 10MΩ 1nF 10MΩ 10MΩ 0.1µF 360kΩ 4.7µF TO DIGITAL INTERFACE 4.7µF RLD SDN 1.8V 50kΩ PD3-4 PD1-2 PDC LEDX1 VLED VREF 1µF AGND DGND LGND 0.1µF 0.1µF DVDD AVDD 1.8V 10kΩ 10kΩ 1.8V SCL SDA GPIO0 GPIO1 TO DIGITAL INTERFACE ADPD105 Figure 77. Synchronous ECG and PPG Measurement Circuit |
|
|
链接网址 |
| ALLDATASHEET是否为您带来帮助? [ DONATE ] |
关于 Alldatasheet | 广告服务 | 联系我们 | 隐私政策 | 数据表链接 | 链接交换 | 制造商名单 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |