12 / 20 page ![]() Type SE35 + S077 12 | 20 Visit product website 5.2. Transmitter Type SE35 mounted in an Inline sensor-fitting Type S077 Note: Dimensions in mm, unless otherwise stated DN H 15 126 25 135 40 147 50 157 80 207 100 223 Threaded connection Flanged connection DN 15 DN 25 DN 40 DN 50 DN 80 DN 25 DN 40 DN 50 DN 80 DN 100 6. Product installation 6.1. Installation notes Note: The device is not suitable for use in gaseous media and steam. The sensor fitting can be installed in any orientation as long as the rotor shafts are always in a horizontal plane. Correct Incorrect The following installation conditions must also be observed: • The pipe always has to be filled with fluid at all times near the device. • The pipe design must be such that no air bubbles or cavitation can form within the medium near the device at any time. • We recommend the installation of a 250 µm strainer as close as possible to the inlet side of the meter, to prevent damage from particles, • Air purges can damage the appliance and should therefore be avoided. |
|
|
链接网址 |
| ALLDATASHEET是否为您带来帮助? [ DONATE ] |
关于 Alldatasheet | 广告服务 | 联系我们 | 隐私政策 | 数据表链接 | 链接交换 | 制造商名单 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|