33 / 36 page ![]() TROUBLESHOOTING cont. Following are troubleshooting displays which could appear on the Model 300/301. Display Cause Solution FACT PARAM FAIL The factory parameters dealing Press ENT to continue. This loads the default values for the with level measurement are corrupt. factory parameters. TAG LINE LOSS The tag line characters are corrupt. Press ENT to continue. This loads the default tag line into long-term memory. If you are using a custom tag line, the tag line must be re-entered. SYS PARAM FAIL The main parameters, which store Press ENT to continue. This loads default values for the main range, span, and communications, parameters. Verify all of the items in the System Config menu. etc. are corrupt. STRAP TABLE FAIL There is an error in the custom Press ENT to continue; the table is cleared. If a custom table is not table (linear or spline) entry. being used, then this error can be ignored. If you are using a custom table, the table has to be re-entered. FV PARAM FAIL The flow or volume parameters Reconfigure flow and volume settings. are corrupt. LOOP PRESET FAIL The loop presets are corrupt. The loop will have to be recalibrated in the Advance Config menu under Adjust Loop Points. RELAY PARAM FAIL The parameters for Relays Press ENT to continue. The relays are disabled, and the settings are are corrupt. set to default values. Reconfigure the relays from the I/O Config menu. LOOP PARAM FAIL The 4–20 mA loop parameters Press ENT to continue. The loop is disabled and the settings are set are corrupt. to default values. Reconfigure the 4-20 mA from the I/O Config menu. TRANSDUCER FAIL The unit was not able to detect a Ensure that the transducer is connected properly to the unit. Try NO TRANSDUCER transducer connected or was not replacing the transducer. Return the unit for repair. OR able to determine the transducer’s BAD TRANSDUCER frequency. ECHO LOSS The unit is not able to detect a Ensure that the level to be detected is within the range and span of return signal from the transducer. the unit. Check the application for foam, obstructions, or heavy vapors. RINGDOWN The transducer is ringing long Try decreasing the span to increase the dead band (range minus span) enough to ring into the level span of of the unit. Try replacing the transducer because the current trans- the unit. ducer may have a problem that causes it to ring an extra long time. TEMPERATURE FAIL The signal from the temperature Check to see that the transducer wires are connected properly. Ensure transducer in the transducer is out that the transducer wires are not shorted or open. of range. Either a short or open lead detected. INVALID PASSWORD The password entered is not valid. Retry the previous operation and enter the proper password. LOOP OUT RANGE The calculated value for the loop If the level is within a normal operating range, adjust the loop output is greater than 20 mA or parameters to allow proper tracking of the level. less than 4 mA. The loop output will be in its fail-safe mode. 33 |
|
链接网址 |
| ALLDATASHEET是否为您带来帮助? [ DONATE ] |
关于 Alldatasheet | 广告服务 | 联系我们 | 隐私政策 | 数据表链接 | 链接交换 | 制造商名单 All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |